Learn · free, no account
Quality-system template outlines
These are the 22 starting points a workspace gets in Regulatory Copilot: the sections each standard actually asks for, deliberately an outline rather than finished prose. Use them freely. In the app, each one becomes a controlled document you can draft from your device's records, route for approval and sign.
Starting points, not compliant documents. Customise each for your device and verify it before use; nothing here is a compliance claim until you have reviewed it, filled in the device-specific detail and approved it.
Quality Manual
ISO 13485 §4.2.2
- Scope of the quality management system, and any clause you exclude (with the justification).
- The processes of the QMS, their sequence and how they interact.
- Reference to the documented procedures that support each process.
- The structure of the documentation, including the medical device file per §4.2.3.
Document & Record Control Procedure
ISO 13485 §4.2.4
- How documents are reviewed and approved before issue, and who approves what.
- How changes are reviewed and approved (by the original approver unless you say otherwise).
- How the current revision is identified and made available at the point of use.
- Retention periods, and how obsolete documents are prevented from unintended use.
CAPA Procedure
ISO 13485 §8.5.2
- Sources of nonconformity that can open a CAPA, and how they are triaged.
- Investigation and root-cause determination, with the depth scaled to the risk.
- Action planning, including whether the action needs design or process change.
- Verification that the action did not adversely affect the device, and the effectiveness check that closes the record.
Complaint Handling & MDR Procedure
21 CFR 803
- How complaints are received, logged verbatim, and acknowledged.
- The reportability assessment: death, serious injury, or malfunction likely to cause one, and who decides.
- MDR timelines (30 days; 5 days where remedial action is required) and the filing route.
- When a complaint escalates into a CAPA, and how the two records stay linked.
Management Review Procedure
ISO 13485 §5.6
- Review frequency, attendees and the documented agenda.
- Inputs: audits, feedback and complaints, process and product conformity, CAPA status, regulatory changes.
- Outputs: improvement decisions, resource decisions, and changes to the QMS.
- How actions are assigned, tracked and closed before the next review.
Design & Development Plan
ISO 13485 §7.3.2
- What is being designed, and the development stages with their entry and exit criteria.
- Responsibilities and authorities, including who performs independent review.
- The verification, validation and design-transfer activities planned at each stage.
- How the plan is updated as the design evolves, and where the records live.
Risk Management Plan
ISO 14971
- Scope of the risk management activities and the lifecycle phases they cover.
- Risk acceptability criteria, including the severity and probability scales used.
- Responsibilities, review requirements, and verification of risk-control implementation and effectiveness.
- How production and post-production information feeds back into the risk file.
Software Development Plan
IEC 62304
- The software safety classification (A / B / C) and the rationale for it.
- Lifecycle model, deliverables and the traceability approach from requirement to test.
- Configuration management, problem resolution and change control.
- SOUP inventory, and how SOUP anomaly lists are reviewed.
Master Test Protocol
Design V&V
- Purpose, the requirements under test and the acceptance criteria for each.
- Test articles, equipment and calibration state.
- Sample size with its statistical rationale, and the method step by step.
- Deviation handling, and what the report must contain to close the protocol.
Software Requirements Specification
IEC 62304
- Functional and performance requirements, each stated so it can be verified.
- Inputs and outputs, interfaces, and alarm or error behaviour.
- Requirements arising from the risk file, including risk-control measures implemented in software.
- Traceability to system requirements and to the verification that covers each item.
Usability Engineering / HF Plan
IEC 62366-1
- Intended users, use environments and the user interface in scope.
- Use-related hazards and the critical tasks derived from them.
- Formative evaluation activities and how findings feed the design.
- The summative (validation) study design, including participant numbers and acceptance criteria.
510(k) Cover Letter
eSTAR
- Submitter and correspondent details, and the device trade name.
- The submission type, the predicate cited, and the regulation and product code.
- A short statement of the purpose of the submission and any prior interactions (Q-Sub numbers).
- The list of what is enclosed, matching the eSTAR sections.
Substantial Equivalence Comparison
510(k)
- Side-by-side table: intended use, indications, technology, materials, performance.
- For each difference, whether it raises a new question of safety or effectiveness.
- The performance data that supports each difference.
- The substantial-equivalence conclusion, stated as your argument, not as a foregone outcome.
Indications for Use (FDA 3881)
FDA form
- Device trade name and the indications-for-use statement, word for word.
- Prescription use or over-the-counter use: tick exactly one.
- Keep this statement identical everywhere it appears in the submission and in the labeling.
Clinical Evaluation Plan
PMA · 21 CFR 814
- The clinical question, the endpoints and the benefit-risk framing.
- Study design, population, sites, and the statistical analysis plan.
- IDE status and the monitoring and data-management arrangements.
- How the clinical evidence maps to the design validation element of the DHF.
Manufacturing & Process Validation Plan
21 CFR 820.70
- The processes whose output cannot be fully verified by inspection, and why.
- IQ / OQ / PQ approach, with acceptance criteria and sample sizes.
- Process monitoring, control limits and revalidation triggers.
- Readiness for the pre-approval inspection: records, training and equipment qualification.
Establishment Registration & Listing SOP
21 CFR 807
- Who registers, when, and the annual registration window (1 October – 31 December).
- Device listing entries and how they are kept current as products change.
- The establishment-registration fee and how payment is evidenced.
- Who holds the owner/operator account, and how official correspondent details are maintained.
Labeling & UDI Procedure
21 CFR 801 / 830
- Required label content and the labeling review and approval route.
- UDI construction: device identifier, production identifiers, and the issuing agency.
- GUDID submission and how records are kept current.
- Label change control, including when a change needs a new UDI-DI.
Post-Market Surveillance Plan
ISO 13485 §8.2.1 · 21 CFR 820.198, 803
- Scope: the devices, markets and data sources the plan covers.
- Sources of post-market data: complaints, MDRs, service records, MAUDE and recalls for comparable devices, literature.
- How often each source is reviewed, by whom, and the trend criteria that trigger action.
- Links to risk management (ISO 14971 production and post-production information) and to CAPA.
- Reporting: MDR timelines (21 CFR 803) and the periodic summary this plan feeds.
Post-Market Surveillance Report
ISO 13485 §8.2.1
- The period covered and the data sources reviewed.
- Complaints and adverse events: counts, trends and the threshold each was judged against.
- Comparable devices: MAUDE and recall signals for the product code.
- Conclusions for risk management and any CAPA opened.
Biocompatibility Evaluation Plan
ISO 10993-1 · FDA guidance on ISO 10993-1
- Each patient-contacting component: material, contact type and duration category.
- The biological endpoints ISO 10993-1 calls for, given those categories.
- For each endpoint: existing data, a rationale, or the test planned (with the standard and lab).
- Chemical characterisation (ISO 10993-18) and toxicological risk assessment where used.
Software Validation Plan
IEC 62304 · FDA guidance on software in a premarket submission
- The software, its safety class (IEC 62304) and its documentation level.
- Verification activities: unit, integration and system testing, with acceptance criteria.
- Validation against user needs, and how anomalies are evaluated and resolved.
- Configuration management, SOUP, and cybersecurity testing where applicable.